I have claimed that cells designed with PLOP are better than older cell designs. In order to support that claim I present here some images of PLOP evaluations of different cells.
|
Remember that these deformations are surface deformation. Light arriving at your eye traverses this distance twice. 1/17 wave surface deformation becomes 1/(8.5) wave path difference at the image. |
|
|
It doesn't require much thought to see that you would rather have your mirror in the second cell than the first. This is especially so when you realize that the central low zone (red) is obscured by the diagonal mirror so that the light reaching your eye comes only from the green and blue areas.
The next seven images were also produced by PLOP. The colors encode, from bright blue to bright red, a range of 5 x 10-5mm or1/10 wave. All deformations are listed as surface deformations, not wavefront errors, which would be twice as large. The best fit paraboloid has been subtracted from each plot, so that these plots show not the total deflection, but the deviation from a parbolic curve.
All seven mirrors represented have 1:6 thickness ratios. All are f/6 and each was calculated assuming an appropriately sized secondary.
This plot shows an 8-inch diameter mirror supported by three points located at 0.393 r. This is the optimum spacing as calculated by PLOP. The P-V deformation is 1/75 wave. The RMS deformation is 1/365 wave.
This plot shows an 8-inch diameter mirror supported by three points located at 0.7071 r. This is the spacing recommended by Hindle. The P-V deformation is 1/30 wave. The RMS deformation is 1/135 wave.
This plot shows an 8-inch diameter mirror supported by three points located at 0.95 r. This is the spacing recommended by Texerau. The P-V deformation is 1/13 wave. The RMS deformation is 1/74 wave.
This plot shows a 10-inch diameter mirror supported by three points located at 0.396 r. This is the optimum spacing as calculated by PLOP. The P-V deformation is 1/48 wave. The RMS deformation is 1/233 wave.
This plot shows a 10-inch diameter mirror supported by three points located at 0.7071 r. This is the spacing recommended by Hindle. The P-V deformation is 1/19 wave. The RMS deformation is 1/86 wave.
This plot shows a 12.5-inch diameter mirror supported by three points located at 0.401 r. This is the optimum spacing as calculated by PLOP. The P-V deformation is 1/30 wave. The RMS deformation is 1/149 wave.
This plot shows a 12.5-inch diameter mirror supported by three points located at 0.7071 r. This is the spacing recommended by Hindle. The P-V deformation is 1/12 wave. The RMS deformation is 1/56 wave.
© 2002 Mark D. Holm